Product Fixes - PF1021
Release Date: 1999/03/31 Affected Model(s): 9407 Affected Product(s): LVDAM-COM 1.01 Brief Description of the Problem: When using the 9407 Virtual Test Equipment Interface with the Spectrum Analyzer, the spectral information for the 20 kHz/Div and 50 kHz/Div spans might have a small frequency offset. Full Description of the Problem: The central frequencies of the filters in the Spectrum Analyzer PCB inside the 9407 module need to be calibrated. A subroutine is missing in the initialization of the module and in the autocalibration of the Spectrum Analyzer program. How to Fix: Replace files TAIVTH00.EXE and TAIVTH01.EXE in folder \LVDAM\COM1_01\. Those files are available by downloading and running the self-extracting file PF1021.EXE. |
|||||||||||||
|